A flexible 0. 4-mm-dia pH probe potentially suitable for physiological use has been developed. It is based on the concept of utilizing two single plastic fiber optic strands to illuminate and remotely sense the color change of a dye indicator contained within an acutely implanted sealed cellulosic hollow fiber permeable to hydrogen ions. A supporting electronic module provides tungsten filament illumination, light sensing with a photodiode/operational amplifier, analog and digital circuitry to provide appropriate signal averaging and processing, and a mechanical assembly to enable the optical density measurements to be made both at 560 nm and, for normalization purposes, in the red. Over the physiological pH range from 7.0 to 7.4, the fiber optic probe agrees with a standard glass pH electrode to within 0.01 pH units in buffer solutions, to within 0.017 pH units in heparinized dog blood in vitro, and it has performed successfully while implanted in the jugular vein of a sheep.
Skip Nav Destination
Article navigation
Research Papers
A Miniature Fiber Optic pH Sensor for Physiological Use
S. R. Goldstein,
S. R. Goldstein
Mechanical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
Search for other works by this author on:
J. I. Peterson,
J. I. Peterson
Chemical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
Search for other works by this author on:
R. V. Fitzgerald
R. V. Fitzgerald
Chemical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
Search for other works by this author on:
S. R. Goldstein
Mechanical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
J. I. Peterson
Chemical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
R. V. Fitzgerald
Chemical Engineering Section, Biomedical Engineering and Instrumentation Branch, Division of Research Services, National Institute of Health, Bethesda, Md. 20014
J Biomech Eng. May 1980, 102(2): 141-146 (6 pages)
Published Online: May 1, 1980
Article history
Received:
October 2, 1979
Revised:
December 23, 1979
Online:
June 15, 2009
Citation
Goldstein, S. R., Peterson, J. I., and Fitzgerald, R. V. (May 1, 1980). "A Miniature Fiber Optic pH Sensor for Physiological Use." ASME. J Biomech Eng. May 1980; 102(2): 141–146. https://doi.org/10.1115/1.3138210
Download citation file:
Get Email Alerts
Analysis of Transient Cutting Forces in Cortical Bone During Ultrasonically Assisted Cutting
J Biomech Eng (June 2025)
Related Articles
Pre-Clinical Evaluation of Direct Current Ablation for the Treatment of Benign Prostatic Hyperplasia
J. Med. Devices (June,2009)
Three-Dimensional Computational Fluid Dynamics Modeling and Validation of Ion Current Sensor in a Gen-Set Diesel Engine Using Chemical Kinetic Mechanism
J. Eng. Gas Turbines Power (October,2017)
Design and Functional Testing of a Novel Blood Pulse Wave Velocity Sensor
J. Med. Devices (March,2018)
Sensing and Control of Double-Sided Arc Welding Process
J. Manuf. Sci. Eng (August,2002)
Related Proceedings Papers
Related Chapters
Fabrication of Silicon Carbide Photodiodes with Enhanced High-Temperature Performance by Electrochemical Anodization Technology
International Conference on Computer and Electrical Engineering 4th (ICCEE 2011)
Two Advanced Methods
Applications of Mathematical Heat Transfer and Fluid Flow Models in Engineering and Medicine
Introduction
Mechanical Blood Trauma in Circulatory-Assist Devices