In this paper, an analysis of the dynamics in the closing phase of the occluder of a mechanical monoleaflet heart valve prosthesis is presented. The dynamic analysis of the fluid in the vicinity of the occluder was based on the control volume approach. The backflow velocity of the fluid was computed by applying the continuity and momentum equations in the unsteady state. By considering the fluid pressure and gravity as external forces acting on the occluder, the moment equilibrium on the occluder was employed to analyze the motion of the occluder during closing and the force of impact between the occluder and the guiding struts. The computed magnitudes of the occluder tip velocities, as well as the backflow of the fluid during the closing phase using this model, were in agreement with previously reported experimental measurements. The maximum impact force between the occluder and guiding struts of 140–280 N was determined to occur during the initial impact for a duration of 35–45 μus. The results of such model studies may be extended for the analysis of the endurance limit of the valve prostheses as well as to determine the mechanical stresses on the formed elements and the incipience of cavitation bubbles during the closing phase of the valve function.
Skip Nav Destination
Article navigation
November 1994
Research Papers
Transient Behavior Analysis of a Mechanical Monoleaflet Heart Valve Prosthesis in the Closing Phase
Gill-Jeong Cheon,
Gill-Jeong Cheon
Department of Mechanical Engineering, Won-Kwang University, Iri City, Jeon-Buk, 570-749 Korea
Search for other works by this author on:
K. B. Chandran
K. B. Chandran
Department of Biomedical Engineering, The University of Iowa, Iowa City, IA 52242
Search for other works by this author on:
Gill-Jeong Cheon
Department of Mechanical Engineering, Won-Kwang University, Iri City, Jeon-Buk, 570-749 Korea
K. B. Chandran
Department of Biomedical Engineering, The University of Iowa, Iowa City, IA 52242
J Biomech Eng. Nov 1994, 116(4): 452-459 (8 pages)
Published Online: November 1, 1994
Article history
Received:
October 25, 1992
Revised:
January 10, 1994
Online:
March 17, 2008
Citation
Cheon, G., and Chandran, K. B. (November 1, 1994). "Transient Behavior Analysis of a Mechanical Monoleaflet Heart Valve Prosthesis in the Closing Phase." ASME. J Biomech Eng. November 1994; 116(4): 452–459. https://doi.org/10.1115/1.2895796
Download citation file:
Get Email Alerts
Related Articles
Cavitation Dynamics of Medtronic Hall Mechanical Heart Valve Prosthesis: Fluid Squeezing Effect
J Biomech Eng (February,1996)
Observation and Quantification of Gas Bubble Formation on a Mechanical Heart Valve
J Biomech Eng (August,2000)
Transient Response of Submerged Plates Subject to Underwater Shock Loading: An Analytical Perspective
J. Appl. Mech (July,2008)
A Dynamic Model of a Contacting Mechanical Seal for Down-Hole Tools
J. Tribol (July,2003)
Related Proceedings Papers
Related Chapters
Dynamic Behavior of Pumping Systems
Pipeline Pumping and Compression Systems: A Practical Approach
Dynamic Behavior of Pumping Systems
Pipeline Pumping and Compression Systems: A Practical Approach, Second Edition
Dynamic Behavior of Pumping Systems
Pipeline Pumping and Compression System: A Practical Approach, Third Edition