A significant potential for plastic hypodermic needles exists as an alternative to current steel needles. This paper presents the design and testing of one type of plastic hypodermic needle. The buckling and penetration characteristics of the needles were modeled and analyzed analytically and by finite element analyses. Experimental penetration tests using plastic and steel control hypodermic needles with skin mimics, specifically polyurethane film and pig skin, were performed to determine penetration and friction forces. Lubricated plastic needles achieved successful penetrations in 25–75% of the tests in both polyurethane film and pig skin with penetration forces ranging from 7 N to 10 N. Penetration tests into butyl rubber stoppers also were conducted with lubricated plastic needles penetrating in 75% of the tests with an average penetration force of 8.3 N. Various lubricants, including silicone oil and a medical grade silicone dispersion, were also studied. In addition, the needles underwent successful perpendicular bending tests and cannula stiffness tests. Finally, fluid flow tests determined fluid flow rates through the needles. Experimental results were compared with each other and to finite element analyses and are discussed.
Skip Nav Destination
Article navigation
December 2009
Research Papers
Characterization of Plastic Hypodermic Needles
Eric Busillo,
Eric Busillo
School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332-0405
Search for other works by this author on:
Jonathan S. Colton
Jonathan S. Colton
School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332-0405
Search for other works by this author on:
Eric Busillo
School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332-0405
Jonathan S. Colton
School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332-0405J. Med. Devices. Dec 2009, 3(4): 041004 (8 pages)
Published Online: November 20, 2009
Article history
Received:
November 14, 2008
Revised:
October 1, 2009
Online:
November 20, 2009
Published:
November 20, 2009
Citation
Busillo, E., and Colton, J. S. (November 20, 2009). "Characterization of Plastic Hypodermic Needles." ASME. J. Med. Devices. December 2009; 3(4): 041004. https://doi.org/10.1115/1.4000452
Download citation file:
Get Email Alerts
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
A Portable Ultrasound-Guided Surgical System for Breast and Thyroid Surgery
J. Med. Devices (June 2025)
The History, Innovation, and Standards of Intravenous Access Ports and Devices
J. Med. Devices (June 2025)
Related Articles
A Minimally Invasive Device for Continuous Glucose Monitoring in Infants
J. Med. Devices (June,2008)
Can Hypodermic Needle Injections Be Successfully Simulated and Analyzed in the Laboratory?
J. Med. Devices (June,2011)
Failure of Locally Buckled Pipelines
J. Pressure Vessel Technol (May,2007)
Ultimate Bending Capacity and Buckling of Pressurized 90 deg Steel Elbows
J. Pressure Vessel Technol (August,2006)
Related Proceedings Papers
Related Chapters
Section VIII: Division 2–Alternative Rules
Companion Guide to the ASME Boiler & Pressure Vessel Codes, Volume 2, Sixth Edition
Subsection NE — Class MC Components
Companion Guide to the ASME Boiler and Pressure Vessel Code, Volume 1, Third Edition
Subsection NE—Class MC Components
Companion Guide to the ASME Boiler & Pressure Vessel Code, Volume 1, Second Edition