Renal disease is epidemic in the United States with approximately 8 × 106 people having chronic kidney disease. Renal biopsies are widely used to provide essential diagnostic information to physicians. However, the risk of bleeding complications possibly leading to life-threatening situations results in the contra-indication of biopsy in certain patient populations. Safer renal biopsies will allow more accurate diagnosis and better management of this epidemic health problem. We report the preclinical testing of a novel biopsy device called the therapeutic injection system (TIS). The device introduces a third stage to the standard two-stage side-cut percutaneous biopsy process. The third stage is designed to reduce bleeding complications by injecting a hemostatic plug at the time of biopsy. Laboratory evaluation and preliminary in vivo animal testing using an anticoagulated porcine model of the TIS and Bard Monopty® (Bard Medical, Covington, GA) control device were performed. The hemostatic material Gelfoam® (Pfizer, Brussels, Belgium) was selected as the active material comprising the hemostatic plugs. The performance of two composite plugs, one composed of polyvinyl alcohol (PVA) combined in 2:1 and 12:1 ratios with the hemostatic material, and one plug composed of 100 hemostatic material were tested. Stroke sequence and hemostatic plug deployment were verified by sequential firing of the TIS biopsy needle into clear gelatin and ex vivo bovine kidney specimens. In vivo trials with porcine specimens revealed a significant reduction in blood loss (8.1 3.9 ml, control versus 1.9 1.6 ml, 12:1 PVA/hemostatic, TIS, = 0.01, = 6). The 100 hemostatic plug showed a substantial and immediate reduction in blood loss (9.2 ml, control versus 0.0 ml, TIS, = 1). The prototype device was shown to work repeatedly and reliably in laboratory trials. Initial results show promise in this approach to control post biopsy bleeding. This solution maintains the simplicity and directness of the percutaneous approach, while not significantly changing the standard percutaneous biopsy procedure.
Skip Nav Destination
University of Michigan,
Department of Anesthesiology,
1500 E. Medical Center Dr., 1H247 UH,
Ann Arbor, MI 48109
e-mail: ghkruger@umich.edu
University of Michigan,
Ann Arbor, MI 48109;
VA Medical Center,
2215 Fuller Road,
Ann Arbor, MI 48105
e-mail: weitzel@umich.edu
Article navigation
March 2013
Research-Article
Performance of Biopsy Needle With Therapeutic Injection System to Prevent Bleeding Complications
Roscoe L. Warner,
Roscoe L. Warner
e-mail: roscoew@umich.edu
University of Michigan,
Ann Arbor, MI 48109
Department of Pathology
,University of Michigan,
1301 Catherine, 7520 MSRBI
,Ann Arbor, MI 48109
Search for other works by this author on:
Grant H. Kruger,
University of Michigan,
Department of Anesthesiology,
1500 E. Medical Center Dr., 1H247 UH,
Ann Arbor, MI 48109
e-mail: ghkruger@umich.edu
Grant H. Kruger
Department of Mechanical Engineering
,University of Michigan,
2350 Hayward, 1031 H.H. Dow
;Department of Anesthesiology,
1500 E. Medical Center Dr., 1H247 UH,
Ann Arbor, MI 48109
e-mail: ghkruger@umich.edu
Search for other works by this author on:
William F. Weitzel
University of Michigan,
Ann Arbor, MI 48109;
VA Medical Center,
2215 Fuller Road,
Ann Arbor, MI 48105
e-mail: weitzel@umich.edu
William F. Weitzel
1
Nephrology Division
,University of Michigan,
102 Observatory, 312 Simpson
,Ann Arbor, MI 48109;
VA Medical Center,
2215 Fuller Road,
Ann Arbor, MI 48105
e-mail: weitzel@umich.edu
1Corresponding author. Disclosure: William Weitzel has ownership interest in Arbor Ultrasound Technologies, LLC.
Search for other works by this author on:
Philip Wong
Kent J. Johnson
e-mail: kjjkjj@umich.edu
Roscoe L. Warner
e-mail: roscoew@umich.edu
University of Michigan,
Ann Arbor, MI 48109
Department of Pathology
,University of Michigan,
1301 Catherine, 7520 MSRBI
,Ann Arbor, MI 48109
Scott I. Merz
Grant H. Kruger
Department of Mechanical Engineering
,University of Michigan,
2350 Hayward, 1031 H.H. Dow
;Department of Anesthesiology,
1500 E. Medical Center Dr., 1H247 UH,
Ann Arbor, MI 48109
e-mail: ghkruger@umich.edu
William F. Weitzel
Nephrology Division
,University of Michigan,
102 Observatory, 312 Simpson
,Ann Arbor, MI 48109;
VA Medical Center,
2215 Fuller Road,
Ann Arbor, MI 48105
e-mail: weitzel@umich.edu
1Corresponding author. Disclosure: William Weitzel has ownership interest in Arbor Ultrasound Technologies, LLC.
Manuscript received January 16, 2012; final manuscript received November 6, 2012; published online February 4, 2013. Assoc. Editor: Foster B. Stulen.
J. Med. Devices. Mar 2013, 7(1): 011002 (7 pages)
Published Online: February 4, 2013
Article history
Received:
January 16, 2012
Revision Received:
November 6, 2012
Citation
Wong, P., Johnson, K. J., Warner, R. L., Merz, S. I., Kruger, G. H., and Weitzel, W. F. (February 4, 2013). "Performance of Biopsy Needle With Therapeutic Injection System to Prevent Bleeding Complications." ASME. J. Med. Devices. March 2013; 7(1): 011002. https://doi.org/10.1115/1.4023274
Download citation file:
Get Email Alerts
Cited By
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
A Portable Ultrasound-Guided Surgical System for Breast and Thyroid Surgery
J. Med. Devices (June 2025)
The History, Innovation, and Standards of Intravenous Access Ports and Devices
J. Med. Devices (June 2025)
Related Articles
Computational Modeling of Blood Hydrodynamics and Blockage Formation Phenomena in the Human Cardiovascular System
J. Med. Devices (June,2010)
Pump Design for a Portable Renal Replacement System
J. Med. Devices (September,2011)
Noninvasive Blood Perfusion Measurements of an Isolated Rat Liver and an Anesthetized Rat Kidney
J Biomech Eng (December,2008)
Blood Perfusion Measurements by the Analysis of the Heated Thermocouple Probe’s Temperature Transients
J Biomech Eng (February,1979)
Related Proceedings Papers
Related Chapters
Investigation of the Energy Shielding of Kidney Stones by Cavitation Bubble Clouds during Burst Wave Lithotripsy
Proceedings of the 10th International Symposium on Cavitation (CAV2018)
Modeling, Simulating, and Synthesizing of Simplified Renal Mechanism
Intelligent Engineering Systems Through Artificial Neural Networks, Volume 17
Modeling, Simulating, and Synthesizing of Simplified Kidney Function
Intelligent Engineering Systems through Artificial Neural Networks Volume 18