We tested the hypothesis that a slight modification in fabrication from the Driver to the Integrity stent (Medtronic) results in nearly equivalent distributions of wall shear stress (WSS) and mean exposure time (MET), reflective of flow stagnation, and that these differences are considerably less than the Multi-Link Vision (Abbott Vascular) or BX Velocity (Cordis) bare metal stents when evaluated by computational fluid dynamics (CFD). Arteries were modeled as idealized straight rigid vessels without lesions. Two vessel diameters (2.25 and 3.0 mm) were studied for each stent and 2.75 mm diameter Integrity stents were also modeled to quantify the impact from best- and worst-case orientations of the stent struts relative to the primary blood flow direction. All stents were 18 mm in length and over-deployed by 10%. The results indicated that, regardless of diameter, the BX Velocity stents had the greatest percentage of the vessel exposed to adverse WSS followed by the Vision, Integrity, and Driver stents. In general, when strut thickness and stent:lumen ratio are similar, the orientation of struts is a determining factor for deleterious flow patterns. For a given stent, the number of struts was a larger determinant of adverse WSS and MET than strut orientation, suggesting that favorable blood flow patterns can be achieved by limiting struts to those providing adequate scaffolding. In conclusion, the Driver and Integrity stents both limit their number of linkages to those which provide adequate scaffolding while also maintaining similar strut thickness and stent:lumen ratios. The Integrity stent also imparts a slight helical velocity component. The modest difference in the fabrication approach between the Driver and Integrity stents is, therefore, not hemodynamically substantial in this idealized analysis, particularly relative to potentially adverse flow conditions introduced by the other stents modeled. This data was used in conjunction with associated regulatory filings and submitted to the FDA as part of the documents facilitating the recent approval for sale of the Resolute Integrity stent in the United States.
Skip Nav Destination
Santa Rosa, CA 95403
Article navigation
March 2013
Research-Article
Computational Fluid Dynamics Evaluation of Equivalency in Hemodynamic Alterations Between Driver, Integrity, and Similar Stents Implanted Into an Idealized Coronary Artery
Timothy J. Gundert,
Timothy J. Gundert
Department of Biomedical Engineering,
Milwaukee, WI 53233
Marquette University
,1515 West Wisconsin Avenue
,Milwaukee, WI 53233
Search for other works by this author on:
Ronak J. Dholakia,
Ronak J. Dholakia
Department of Neurological Surgery,
Stony Brook University Medical Center
,Stony Brook, NY 11794
Search for other works by this author on:
Dennis McMahon,
Santa Rosa, CA 95403
Dennis McMahon
Medtronic CardioVascular
,3576 Unocal Place
,Santa Rosa, CA 95403
Search for other works by this author on:
John F. LaDisa, Jr.
John F. LaDisa, Jr.
1
Department of Biomedical Engineering,
Milwaukee, WI 53233;
Division of Cardiovascular Medicine,
Medical College of Wisconsin,
Milwaukee, WI 53226
e-mail: john.ladisa@marquette.edu
Marquette University
,1515 West Wisconsin Avenue
,Milwaukee, WI 53233;
Department of Medicine
,Division of Cardiovascular Medicine,
Medical College of Wisconsin,
8701 Watertown Plank Road
,Milwaukee, WI 53226
e-mail: john.ladisa@marquette.edu
1Corresponding author.
Search for other works by this author on:
Timothy J. Gundert
Department of Biomedical Engineering,
Milwaukee, WI 53233
Marquette University
,1515 West Wisconsin Avenue
,Milwaukee, WI 53233
Ronak J. Dholakia
Department of Neurological Surgery,
Stony Brook University Medical Center
,Stony Brook, NY 11794
Dennis McMahon
Medtronic CardioVascular
,3576 Unocal Place
,Santa Rosa, CA 95403
John F. LaDisa, Jr.
Department of Biomedical Engineering,
Milwaukee, WI 53233;
Division of Cardiovascular Medicine,
Medical College of Wisconsin,
Milwaukee, WI 53226
e-mail: john.ladisa@marquette.edu
Marquette University
,1515 West Wisconsin Avenue
,Milwaukee, WI 53233;
Department of Medicine
,Division of Cardiovascular Medicine,
Medical College of Wisconsin,
8701 Watertown Plank Road
,Milwaukee, WI 53226
e-mail: john.ladisa@marquette.edu
1Corresponding author.
Manuscript received July 24, 2012; final manuscript received January 2, 2013; published online February 4, 2013. Assoc. Editor: Keefe B. Manning.
J. Med. Devices. Mar 2013, 7(1): 011004 (10 pages)
Published Online: February 4, 2013
Article history
Received:
July 24, 2012
Revision Received:
January 2, 2013
Citation
Gundert, T. J., Dholakia, R. J., McMahon, D., and LaDisa, J. F., , Jr. (February 4, 2013). "Computational Fluid Dynamics Evaluation of Equivalency in Hemodynamic Alterations Between Driver, Integrity, and Similar Stents Implanted Into an Idealized Coronary Artery." ASME. J. Med. Devices. March 2013; 7(1): 011004. https://doi.org/10.1115/1.4023413
Download citation file:
Get Email Alerts
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
A Portable Ultrasound-Guided Surgical System for Breast and Thyroid Surgery
J. Med. Devices (June 2025)
The History, Innovation, and Standards of Intravenous Access Ports and Devices
J. Med. Devices (June 2025)
Related Articles
Hydrodynamic Effects of Compliance Mismatch in Stented Arteries
J Biomech Eng (February,2011)
Impact of Stents and Flow Diverters on Hemodynamics in Idealized Aneurysm Models
J Biomech Eng (July,2011)
Coronary Artery Stenting Affects Wall Shear Stress Topological Skeleton
J Biomech Eng (June,2022)
Influence of the Anatomical Structure on the Hemodynamics of Iliac Vein Stenosis
J Biomech Eng (January,2023)
Related Proceedings Papers
Related Chapters
mDFA Human Empirical Results
Modified Detrended Fluctuation Analysis (mDFA)
Some Examples of Evaluating Structural Integrity by Acoustic Emission Monitoring during Pressure Testing
Acoustic Emission Monitoring of Pressurized Systems
In-Vitro Modeling of the Dynamic Forces in the Femoropopliteal Artery
Fatigue and Fracture of Medical Metallic Materials and Devices: 2nd Volume