Huber needles were designed to penetrate an implantable port septum without producing cores, or shaving off slivers of material from the septum which can cause the port to leak or fail. Needles with manufacturing flaws resulting in sharp rear bevel edges [1] have been undetected in some products. That sharp rear bevel edge (sharp anti-coring heel) is the major factor in producing cores (i.e. plugs of septum material) during the septum perforation. When a core is produced the small piece of silicone (a core) can be flushed from the needle into the port, possibly entering the blood vessel. This can lead to adverse health events. In recent years about two million Huber needles were voluntarily recalled from the market due to the possibility of coring. There is no standardized coring test for Huber needles. Some manufacturers sell their ports and needles as a kit while a majority of...
Skip Nav Destination
U.S. Food and Drug Administration,
Article navigation
June 2013
Technical Briefs
Development of a Coring Test for Non-Coring Huber Needles
Jon P. Casamento
U.S. Food and Drug Administration,
Jon P. Casamento
Center for Devices and Radiological Health
,U.S. Food and Drug Administration,
Silver Spring, MD 20993
Search for other works by this author on:
Jon P. Casamento
Center for Devices and Radiological Health
,U.S. Food and Drug Administration,
Silver Spring, MD 20993
Manuscript received March 15, 2013; final manuscript received April 24, 2013; published online June 12, 2013. Editor: Arthur G. Erdman.
J. Med. Devices. Jun 2013, 7(2): 020930 (2 pages)
Published Online: June 12, 2013
Article history
Received:
March 15, 2013
Revision Received:
April 24, 2013
Citation
Vesnovsky, O., and Casamento, J. P. (June 12, 2013). "Development of a Coring Test for Non-Coring Huber Needles." ASME. J. Med. Devices. June 2013; 7(2): 020930. https://doi.org/10.1115/1.4024306
Download citation file:
50
Views
Get Email Alerts
Cited By
Related Articles
A Testbed for Multilumen Steerable Needle Experiments
J. Med. Devices (June,2010)
Characterization of Plastic Hypodermic Needles
J. Med. Devices (December,2009)
Characterization of Precurved Needles for Use in Distal Tip Manipulation Mechanisms
J. Med. Devices (June,2010)
Modeling of the Plane Needle Cutting Edge Rake and Inclination Angles for Biopsy
J. Manuf. Sci. Eng (October,2010)
Related Proceedings Papers
Related Chapters
On Interfacial Ribbons of ζ- and γ-Hydride Phases Surrounding δ Precipitates in Zircaloy-2
Zirconium in the Nuclear Industry: 20th International Symposium
Antilock-Braking System Using Fuzzy Logic
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
Internal and Near Nozzle Flow Simulations of Gasoline Multi-Hole Injector (ECN Spray G) with Transient Needle Motion
Proceedings of the 10th International Symposium on Cavitation (CAV2018)