Reprocessing has emerged as an attempt to control the cost of single-use bipolar and ultrasound diathermy devices despite limited data on defect rates. This study compares the defect rates, as reported by surgical teams, between original equipment manufacturer (OEM) single-use bipolar and ultrasound diathermy devices and reprocessed (RP) devices. Data were retrospectively collected on 3112 devices over a 7-month period for two types of bipolar and ultrasound diathermy devices. There is a significant difference (p < 0.001) in reported bipolar and ultrasound diathermy device defects between OEM and RP. OEM single-use bipolar and ultrasound diathermy devices were reported to be defective more frequently than RP devices based on reports from the surgical team.
Skip Nav Destination
Article navigation
December 2015
Technical Briefs
A Comparison of the Defect Rate Between Original Equipment Manufacturer and Reprocessed Single-Use Bipolar and Ultrasound Diathermy Devices
Terrence J. Loftus
Terrence J. Loftus
Medical Director Surgical Services and Clinical Resources,
Division of Care Management,
Banner Health,
1441 N. 12th Street,
Phoenix, AZ 85006
e-mail: Terry.Loftus@BannerHealth.com
Division of Care Management,
Banner Health,
1441 N. 12th Street,
Phoenix, AZ 85006
e-mail: Terry.Loftus@BannerHealth.com
Search for other works by this author on:
Terrence J. Loftus
Medical Director Surgical Services and Clinical Resources,
Division of Care Management,
Banner Health,
1441 N. 12th Street,
Phoenix, AZ 85006
e-mail: Terry.Loftus@BannerHealth.com
Division of Care Management,
Banner Health,
1441 N. 12th Street,
Phoenix, AZ 85006
e-mail: Terry.Loftus@BannerHealth.com
Manuscript received July 16, 2014; final manuscript received June 8, 2015; published online August 6, 2015. Assoc. Editor: Rosaire Mongrain.
J. Med. Devices. Dec 2015, 9(4): 044501 (2 pages)
Published Online: August 6, 2015
Article history
Received:
July 16, 2014
Revision Received:
June 8, 2015
Citation
Loftus, T. J. (August 6, 2015). "A Comparison of the Defect Rate Between Original Equipment Manufacturer and Reprocessed Single-Use Bipolar and Ultrasound Diathermy Devices." ASME. J. Med. Devices. December 2015; 9(4): 044501. https://doi.org/10.1115/1.4030858
Download citation file:
Get Email Alerts
Cited By
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
A Portable Ultrasound-Guided Surgical System for Breast and Thyroid Surgery
J. Med. Devices (June 2025)
The History, Innovation, and Standards of Intravenous Access Ports and Devices
J. Med. Devices (June 2025)
Related Articles
Data Communication Pathway for Sensing Guidewire at Proximal Side: A Review
J. Med. Devices (June,2017)
Hydraulic Robotic Surgical Tool Changing Manipulator
J. Med. Devices (March,2017)
Assessing the Importance of Surgeon Hand Anthropometry on the Design of Medical Devices
J. Med. Devices (December,2017)
Attention-Aware Robotic Laparoscope Based on Fuzzy Interpretation of Eye-Gaze Patterns
J. Med. Devices (December,2015)
Related Chapters
Oxygen-Enriched Fires During Surgery of The Head and Neck
Flammability and Sensitivity of Materials in Oxygen-Enriched Atmospheres: Fourth Volume
QRAS Approach to Phased Mission Analysis (PSAM-0444)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)
A Human Reliability-Centered Approach to the Development of Job Aids for Reviewers of Medical Devices That Use Radiological Byproduct Materials (PSAM-0299)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)